Calibration slide for microscopes.
The slide has three horizontal micrometer scales of 100 divisions. The scale division is 0.01mm, 0.05mm, and 0.1mm, respectively.
The calibration slide (stage micrometer) is used to determine the magnification, linear field of view of the microscope, and the division value of the eyepiece scales and reticles.
With a digital camera, the slide is used to calibrate the image analysis software for each objective.
Calibrating the image analysis software
The calibration is performed for each objective that will be used in observations. The calibration slide is used to determine the calibration factor for the image analysis software.
Calibration procedure:
Calibration of the eyepiece scale
The calibration of the eyepiece scale is performed for each objective that will be used in observations. The slide should be placed on the stage and the microscope should be focused on the scale. The eyepiece should be rotated in the eyepiece tube until the scales of a slide and an eyepiece are aligned parallel to each other.
If you use low-magnification objectives, you should calculate how many divisions of the eyepiece scale fit into the slide scale.
If you use medium and high-magnification objectives, you should calculate how many divisions of the slide fit into the scale of the eyepiece.
The MAGUS CS3 calibration slide offers three micrometer scales with different division values. Calibration is more convenient when the slide scale covers the entire eyepiece scale. On the low-magnification objective, a scale with a division of 0.1mm or 0.05mm is used. A scale with 0.01mm division is used on the objectives with magnification above 20x.
Calculation for each objective is carried out according to the formula:
Е=ТL/A, where
E – eyepiece division value
Т – stage division value specified on the stage micrometer
L – number of stage micrometer divisions
А – number of eyepiece scale divisions.
The resulting values of the eyepiece scale division for each objective shall be recorded.
When measuring an object, first you should count the number of eyepiece scale divisions that the object covers. The result is multiplied by the scale value previously calculated for the selected objective. The resulting value will be the true linear value of the object.
Brand | MAGUS |
Warranty | 2 years |
EAN | 5905555017198 |
Package size (LxWxH) | 9x4x1 cm |
Shipping Weight | 0.02 kg |
Material | optical glass |
Number of scale divisions | 100 |
Scale value, mm | 0.01/0.05/0.1 |
Calibration slide size, mm | 75x25x1.1 |